AELAB BPM-200 Continuous Particulate Monitor – Real-Time PM Measurement

The AELAB BPM-200 Continuous Particulate Monitor quantifies ambient particulate matter using beta ray attenuation with a controlled 16.7 L/min flow. With 0.1 μg/m³ resolution, selectable ranges up to 10 mg/m³, and EPA-equivalent methodology, the instrument provides stable, real-time PM2.5/PM10 measurement for air quality networks, research stations, and industrial hygiene programs.

Modles : BPM-200
whatsapp Whatsapp

The AELAB BPM-200 Continuous Particulate Monitor is a beta gauge instrument that measures real-time particulate mass concentration for environmental and industrial air quality programs. Using a fixed 16.7 L/min sample flow and reference film calibration, the BPM-200 quantifies PM with high accuracy and stability across selectable ranges up to 10 mg/m³. With optional PM2.5 cyclone separation, the monitor supports regulatory PM2.5 testing and continuous surveillance in urban networks and research-grade stations.

 

⚙ Design & Construction

  • Integrated vacuum sampling draws air onto a glass fiber filter tape that advances automatically for continuous operation up to 60 days.
  • A sealed measurement path positions the tape between a C14 beta source and a photomultiplier tube (PMT) detector for stable counting.
  • Rigid enclosure with LCD user interface provides status, diagnostics, and averaged results from 1 minute to 365 days.
  • Flow controller achieves 16.7 L/min with ±1% F.S. accuracy and ≤±2%/24 h stability (optional).

 

🔬 Working Principle / Testing Method

The AELAB BPM-200 applies beta ray attenuation gravimetry. As collected particles accumulate on the moving filter tape, they attenuate beta radiation transmitted from the C14 source to the PMT. The reduction in count rate is proportional to deposited mass. The instrument calculates particulate concentration by dividing the measured mass by the precisely metered sample volume, delivering real-time PM values. Installing a cyclone enables PM2.5 size-selective sampling before filtration.

 

💡 Key Advantages

  • EPA-equivalent beta attenuation method for defensible reporting.
  • 0.1 μg/m³ resolution and ≤2% accuracy with low drift.
  • Automatic tape advance for continuous up to 60 days operation.
  • Reference film calibration improves long-term comparability.
  • User-selectable measurement cycles to match site objectives.
  • Comprehensive alarms and continuous self-diagnostics for uptime.

 

🔗 Connectivity & Integration

The BPM-200 integrates with supervisory systems through RS232/RS485/Ethernet and analog outputs (0–5 V, 4–20 mA). Digital I/O and relay contacts support sampler control, status signaling, and external device interlocks. Optional GPRS and serial printer expand field reporting and local record keeping.

 

🧠 Model Insight

The AELAB BPM-200 suits long-term fixed stations requiring unattended operation. Selectable ranges ((0–1) to (0–10) mg/m³) and configurable averaging (1 min to 365 days) allow deployment from urban background monitoring to industrial perimeter surveillance. The C14 source provides stable emission with low maintenance, and the PMT ensures sensitive detection for fine particulate trends.

 

🧪 Testing Applications

  • Ambient PM2.5/PM10 monitoring in air quality networks.
  • Industrial hygiene and perimeter dust compliance tracking.
  • Research on aerosol dynamics, pollution episodes, and source apportionment.
  • Regulatory screening with reference film calibration for auditability.

 

Compliance & Quality Assurance

The AELAB BPM-200 Continuous Particulate Monitor follows US EPA equivalent method practices for beta attenuation mass measurement. Each unit is delivered with calibration documentation and functional verification (flow, tape advance, detector counts). AELAB maintains GLP-style procedures for traceability, supporting consistent data quality in continuous particulate monitoring.

Specifications :

Feature Details
Capacity (0–1)/(0–2)/(0–5)/(0–10) mg/m³ (selectable)
Readability 0.1 μg/m³
Units μg/m³, mg/m³
Platform Size
Display LCD
Interface RS232, RS485, Ethernet
Power (220 ±22) V AC, (50 ±1) Hz
Battery
Calibration Reference film calibration; calibration film reproducibility ≤±2%
Output 2× analog (0–5 V), 2× analog (4–20 mA), 12× digital I/O, 4× relay
Dimensions 310 (H) × 430 (W) × 400 (D) mm
Weight ≈25 kg (without accessories)
Construction Beta source: C14 (active 10 μCi, exemption source); PMT detector; glass fiber filter tape (>20 m roll)
Operating Conditions −5–55 °C; 0–95% RH (non-condensing); 86–116 kPa
Accessories PM2.5 cyclone separator (optional), serial printer (optional)
Feature Details
Capacity (0–1)/(0–2)/(0–5)/(0–10) mg/m³ (selectable)
Readability 0.1 μg/m³
Units μg/m³, mg/m³
Platform Size
Display LCD
Interface RS232, RS485, Ethernet
Power (220 ±22) V AC, (50 ±1) Hz
Battery
Calibration Reference film calibration; calibration film reproducibility ≤±2%
Output 2× analog (0–5 V), 2× analog (4–20 mA), 12× digital I/O, 4× relay
Dimensions 310 (H) × 430 (W) × 400 (D) mm
Weight ≈25 kg (without accessories)
Construction Beta source: C14 (active 10 μCi, exemption source); PMT detector; glass fiber filter tape (>20 m roll)
Operating Conditions −5–55 °C; 0–95% RH (non-condensing); 86–116 kPa
Accessories PM2.5 cyclone separator (optional), serial printer (optional)

Frequently Asked Questions

1What principle does the AELAB BPM-200 use to measure particulate matter?
The BPM-200 uses beta ray attenuation on a glass fiber filter tape and converts the attenuation to PM mass concentration using the metered sample volume.
2What flow rate does the BPM-200 operate at?
The standard flow rate is 16.7 L/min with ±1% full-scale accuracy.
3Can the BPM-200 measure PM2.5?
Yes. Adding the appropriate cyclone provides PM2.5 size-selective sampling before beta attenuation measurement.
4What is the instrument resolution and accuracy?
Resolution is 0.1 μg/m³ with ≤2% accuracy across selectable ranges.
5How long can it operate continuously?
With automatic tape advance, the system supports continuous operation up to 60 days.
6What outputs are available for integration?
RS232, RS485, Ethernet, two analog outputs (0–5 V and 4–20 mA), 12 digital I/O, and four relay outputs.
7Is the method EPA-equivalent?
Yes, the BPM-200 complies with US EPA equivalent method criteria for beta attenuation particulate monitoring.
8What calibration method is used?
Reference film calibration is provided to verify and maintain measurement traceability.

Request Quote: AELAB BPM-200 Continuous Particulate Monitor – Real-Time PM Measurement