The AELAB T8000 Flame Atomic Absorption Spectrometer is designed to determine the content of major, trace, and semi-metal elements in a wide range of substances. The instrument integrates advanced electronic circuitry, automatic wavelength adjustment, and an intuitive multilingual operating interface. Equipped with USB2.0 communication, it enables rapid spectral calibration, lamp control, and automated analytical sequencing with high stability and reproducibility.
Through a PC-based control system, the T8000 allows operators to manage wavelength scanning, peak positioning, spectral bandwidth, lamp current, and atomizer height with precision. Stored reference conditions and analysis parameters can be recalled for consistent, repeatable testing across industrial and laboratory environments.
The AELAB T8000 features a titanium atomizing chamber for corrosion resistance and extended service life. Its fully enclosed optical system includes a Cherney Turner monochromator, high-precision grating (1800 lines/mm), and total reflection concave mirror for enhanced light focusing. A high-efficiency glass atomizer ensures superior atomization and strong universality across different metal elements.
The spectrometer operates on the principle of atomic absorption, where atoms in a flame absorb light at specific wavelengths corresponding to their element type. The instrument measures absorbance variations proportional to concentration. Deuterium lamp and self-absorption correction methods enable accurate background compensation, while the five-gear automatic spectral bandwidth switching supports flexible analytical conditions from 0.1 nm to 1.6 nm.
The system features a USB2.0 port for data transfer and real-time communication. Compatible with AELAB analytical software, it supports quantitative and qualitative analysis, automatic report generation, and GLP-compliant data export.
| Model | Wavelength Range | Atomizer Type | Communication | Baseline Stability |
|---|---|---|---|---|
| T8000 | 190–900 nm | Flame Atomizer (Titanium) | USB2.0 Interface | 0.002 Abs (Static/Dynamic) |
The AELAB T8000 is widely used in geological, metallurgical, environmental, chemical, food, and material science laboratories for quantitative metal analysis. Typical applications include soil and water quality monitoring, alloy composition verification, pharmaceutical testing, and trace contamination studies.
Manufactured under ISO and CE compliance, the AELAB T8000 ensures full traceability and GLP conformity. Each unit undergoes multi-stage calibration verification, optical alignment testing, and stability evaluation before release.
Specifications :
| Parameter | Specification |
|---|---|
| Wavelength Range | 190 nm – 900 nm |
| Monochromator Type | Cherney Turner Type |
| Spectral Bandwidth | 0.1 / 0.2 / 0.4 / 0.7 / 1.6 nm, Five-gear automatic switching |
| Grating Scribed Line | 1800 lines/mm |
| Wavelength Accuracy | ±0.20 nm |
| Wavelength Repeatability | <0.05 nm |
| Resolving Power | ≤0.3 nm |
| Characteristic Concentration (Cu) | <0.025 µg/mL / 1% |
| Detection Limit (Cu) | <0.003 µg/mL |
| Grating Blaze Wavelength | 230 nm |
| Baseline Stability | 0.002 Abs (Static) / 0.002 Abs (Dynamic) |
| Background Deduction Method | Deuterium Lamp + Self-Absorption Correction |
| Precision (Cu) | <0.5% |
| Parameter | Specification |
|---|---|
| Wavelength Range | 190 nm – 900 nm |
| Monochromator Type | Cherney Turner Type |
| Spectral Bandwidth | 0.1 / 0.2 / 0.4 / 0.7 / 1.6 nm, Five-gear automatic switching |
| Grating Scribed Line | 1800 lines/mm |
| Wavelength Accuracy | ±0.20 nm |
| Wavelength Repeatability | <0.05 nm |
| Resolving Power | ≤0.3 nm |
| Characteristic Concentration (Cu) | <0.025 µg/mL / 1% |
| Detection Limit (Cu) | <0.003 µg/mL |
| Grating Blaze Wavelength | 230 nm |
| Baseline Stability | 0.002 Abs (Static) / 0.002 Abs (Dynamic) |
| Background Deduction Method | Deuterium Lamp + Self-Absorption Correction |
| Precision (Cu) | <0.5% |
Download catalog for: