The AELAB VC100 Vacuum Controller serves as a laboratory device for displaying, adjusting, and controlling vacuum levels in a wide range of vacuum applications. It integrates seamlessly with rotary evaporators and vacuum pumps, enabling repeatable distillation and efficient solvent recovery in automated workflows.
Built for precision and long-term reliability, the AELAB VC100 supports both single-point and programmable control modes. Users can configure multi-step vacuum profiles to optimize evaporation, reduce solvent bumping, and maintain stable pressure throughout sensitive analytical and preparative procedures.
The AELAB VC100 features a robust housing with a 5” TFT touch screen for clear visualization of real-time pressure and control settings. All wetted components are constructed from PTFE, PP, silicone, and ceramics, providing strong resistance against organic solvents, acids, alkalis, and moisture. Its 8 mm connector interface supports easy integration with diverse vacuum lines, and the built-in vent function provides safe, controlled pressure equalization.
The VC100 regulates vacuum by continuously monitoring pressure between 1–1000 mbar and modulating pump output based on selected control modes. In programmed mode, up to five independent programs can be configured, each containing five timed steps. This enables precise pressure curves tailored to distillation stages, solvent boiling points, or evaporation kinetics. The internal sensor supports up to 1,500 mbar overload, ensuring stable operation under varying load conditions.
The VC100 connects directly to compatible vacuum pumps (≤400 W at 50 Hz or ≤500 W at 60 Hz) and integrates with rotary evaporators, solvent recovery units, and automated evaporation systems. Its interface supports external pump control, automatic venting, and synchronized operation for GLP-compliant data handling.
| Model | Specifications |
|---|---|
| VC100 | 1–1000 mbar range, 5” TFT display, 600 W power, 189x207x193 mm, 4.0 kg |
The AELAB VC100 is manufactured under strict ISO and CE quality standards, ensuring repeatable performance, chemical resistance, and long-term reliability. All sensors and electronics undergo calibration and verification under AELAB’s GLP-compliant production protocols.
Specifications :
| Specifications | VC100 |
|---|---|
| Vacuum setting range | 1–1000 mbar |
| Vacuum measurement range | 1–1000 mbar |
| Control mode | Single-point control mode, Programmed control mode |
| Control program | Can store up to 5 programs, each containing 5-step control (capable to set the degree of vacuum and time) |
| Setting method | Touch screen setting |
| Display | 5'' TFT |
| Sensor overload pressure | 1,500 mbar |
| Connector diameter | 8 mm |
| Materials in contact with vapor | PTFE, PP, silicone and ceramics |
| Power | 600 W |
| Compatible conditions with vacuum pump | Power of 50 Hz vacuum pump ≤400 W; Power of 60 Hz vacuum pump ≤500 W |
| Overall dimensions W×D×H | 189×207×193 mm |
| Weight | 4.0 kg |
| Voltage, Frequency | 110–240 V, 50 Hz / 60 Hz |
| Operating environment | 10°C~40°C <80% RH |
| Specifications | VC100 |
|---|---|
| Vacuum setting range | 1–1000 mbar |
| Vacuum measurement range | 1–1000 mbar |
| Control mode | Single-point control mode, Programmed control mode |
| Control program | Can store up to 5 programs, each containing 5-step control (capable to set the degree of vacuum and time) |
| Setting method | Touch screen setting |
| Display | 5'' TFT |
| Sensor overload pressure | 1,500 mbar |
| Connector diameter | 8 mm |
| Materials in contact with vapor | PTFE, PP, silicone and ceramics |
| Power | 600 W |
| Compatible conditions with vacuum pump | Power of 50 Hz vacuum pump ≤400 W; Power of 60 Hz vacuum pump ≤500 W |
| Overall dimensions W×D×H | 189×207×193 mm |
| Weight | 4.0 kg |
| Voltage, Frequency | 110–240 V, 50 Hz / 60 Hz |
| Operating environment | 10°C~40°C <80% RH |
No download files available for this product.