The AELAB SYD-3536 Cleveland Open Cup Flash Point Tester is designed and manufactured in accordance with GB/T3536-2008 Petroleum Products – Determination of Flash and Fire Points – Cleveland Open Cup Method and ASTM D92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79 °C and below 400 °C except fuel oils.
The instrument adopts a special heating furnace to ensure test safety. The heating power is continuously adjustable to meet the analytical requirements. The tester is fully self-contained, requiring only connection to coal gas or civil gas for flame application. A fast cooling function enhances operational efficiency in routine flash point measurements.
The AELAB SYD-3536 features a furnace-type heating system that provides uniform heating without a naked flame, ensuring explosion-preventive operation. The power is adjustable from 0 W to 600 W for precise temperature rise control. A sturdy metal housing, protected ignition assembly, and enclosed test chamber maintain consistent thermal conditions and operator safety.
The SYD-3536 operates using the Cleveland Open Cup (COC) method, heating the petroleum sample at a controlled rate while automatically applying a test flame. The instrument monitors the vapor–air mixture above the sample and identifies the flash point when a transient flame appears. The thermometer (−6 to 400 °C, 2 °C scale, ASTM 11C equivalent) provides accurate temperature reference throughout the test procedure.
The SYD-3536 includes a digital display for temperature monitoring and integrates easily into petroleum QC labs. Data can be recorded manually following ASTM D92 and GB/T3536 procedures, ensuring full GLP traceability when used with laboratory LIMS systems.
| Model | Specifications |
|---|---|
| SYD-3536 | AC (220±10%) V, 50Hz; ≤650W; 0–600 W furnace heating; automatic flame applicator; −6–400 °C thermometer; 350×290×350 mm; 5.5 kg |
The AELAB SYD-3536 is suitable for petroleum laboratories, fuel depots, energy companies, lubricants manufacturers, and research facilities. Typical applications include flash point determination of lubricating oils, base oils, solvent oils, hydraulic fluids, insulating oils, and high-flash-point petroleum products.
The instrument complies with GB/T3536-2008 and ASTM D92 requirements. All units undergo AELAB QC processes, ensuring stable performance, verified calibration, and long-term reliability in high-temperature petroleum testing environments.
Specifications :
| 1. Power supply | AC (220±10%) V, 50Hz. |
| 2. Power consumption | ≤650W |
| 3. Heating device | furnace heating, no naked fire, explosion prevented. The power is adjustable from 0W to 600W. |
| 4. Test flame applicator | It applies the test flame automatically. |
| 5. Thermometer | (−6~400)°C. Scale is 2°C. It is the same as thermometer ASTM 11C. |
| 6. Igniting device | (1) Ignition source: coal gas (or civil gas) (2) Nozzle aperture: about 0.8mm |
| 7. Ambient temperature | (15~35)°C, Relative humidity: ≤85% |
| 8. Dimension | 350mm × 290mm × 350mm (thermometer is not included) |
| 9. Net weight | 5.5kg |
| 1. Power supply | AC (220±10%) V, 50Hz. |
| 2. Power consumption | ≤650W |
| 3. Heating device | furnace heating, no naked fire, explosion prevented. The power is adjustable from 0W to 600W. |
| 4. Test flame applicator | It applies the test flame automatically. |
| 5. Thermometer | (−6~400)°C. Scale is 2°C. It is the same as thermometer ASTM 11C. |
| 6. Igniting device | (1) Ignition source: coal gas (or civil gas) (2) Nozzle aperture: about 0.8mm |
| 7. Ambient temperature | (15~35)°C, Relative humidity: ≤85% |
| 8. Dimension | 350mm × 290mm × 350mm (thermometer is not included) |
| 9. Net weight | 5.5kg |
Download catalog for: