The AELAB EDS8800 Precious Metal Analyzer is an X-ray fluorescence spectrum analyzer engineered for quantitative elemental analysis and composition verification in precious metal testing workflows. The EDS8800 supports solid, powder, and liquid sample analysis, delivering fast results in 5 – 60 sec with an analysis range from ppm to 99.99% for direct assay screening and quality control.
Designed for laboratories, jewelry manufacturing, refineries, and industrial inspection, the AELAB EDS8800 Precious Metal Analyzer combines a Si-pin detector, digital multichannel spectrum processing, and a selectable filter system to stabilize excitation conditions across multi-element sample matrices. With 145 ± 5 eV energy resolution and 0.03% accuracy (Au ≥ 96%), the EDS8800 enables repeatable elemental measurement from sodium (Na) to uranium (U).
The AELAB EDS8800 Precious Metal Analyzer integrates a compact XRF measurement chamber with a stable high voltage power supply rated 0 – 50 kV and tube voltage operation from 5 kV – 50 kV. The EDS8800 features a sealed sample chamber sized 310 × 270 × 80 mm, supporting diverse specimen geometries for non-destructive fluorescence spectrum analysis. An integrated HD camera improves sample positioning and repeatability for laboratory and industrial assay workflows.
The AELAB EDS8800 Precious Metal Analyzer performs X-ray fluorescence spectrum analysis by exciting sample atoms using a controlled X-ray tube voltage and measuring emitted characteristic fluorescence energy peaks with a Si-pin detector. The EDS8800 quantifies elemental composition through digital multichannel spectral acquisition and mathematical peak deconvolution, enabling qualitative and quantitative measurement for up to 26 elements simultaneously per test cycle. Qualitative and quantitative analysis software processes peak intensities and converts fluorescence signals into concentration results from ppm to 99.99%.
The AELAB EDS8800 Precious Metal Analyzer is supplied with qualitative and quantitative analysis software to support elemental analysis workflows, spectrum interpretation, and report generation. The software architecture enables repeatable quantitative measurement, data storage, and standardized analytical documentation for laboratory testing, production verification, and industrial QC programs.
The AELAB EDS8800 Precious Metal Analyzer supports XRF fluorescence spectrum analysis for elemental composition measurement in jewelry production, refineries, recycling, and industrial laboratories. The EDS8800 is used to quantify gold purity (Au ≥ 96% reference accuracy condition), verify alloy composition, screen for base metal adulteration, and measure multi-element content across precious metal matrices. Typical testing programs include incoming material inspection, process control validation, finished product certification, recycling sorting, and contamination detection for gold, silver, and alloy components.
With the ability to measure 82 elements and analyze up to 26 elements simultaneously, the AELAB EDS8800 Precious Metal Analyzer is also suitable for elemental screening in industrial quality control, metallurgical verification, and laboratory analytical workflows where rapid, non-destructive composition assessment is required.
The AELAB EDS8800 Precious Metal Analyzer is designed for controlled laboratory and industrial operation under recommended conditions of 5 °C – 38 °C and 15% – 90% relative humidity. The instrument requires stable power input of AC 220 V ± 10 V or AC 110 V ± 10 V at 50/60 Hz and should be installed away from high-power electromagnetic fields and vibration interference sources to ensure measurement stability. AELAB quality assurance protocols emphasize repeatable calibration control, traceable analytical workflow structure, and consistent elemental testing documentation for precious metal assay processes.
For laboratories seeking repeatable XRF elemental verification, the AELAB EDS8800 Precious Metal Analyzer delivers validated fluorescence spectrum analysis performance with compact installation and fast test cycle efficiency.
Specifications :
| Analysis Range | ppm to 99.99% |
| Accuracy | 0.03% (Au ≥ 96%) |
| Sample Form | Solid, powder, liquid |
| Tube Voltage | 5kV ~ 50kV |
| High Voltage Power Supply | 0 ~ 50kV |
| Camera | HD camera |
| Filters | Selectable customized switches |
| Detector Type | Si-pin |
| Resolution | 145 ± 5 eV |
| Multichannel Analyzer | Digital multichannel |
| Sample Chamber Size | 310 × 270 × 80 mm |
| Test Time | 5 ~ 60 sec |
| Measuring Element | 82 elements from sodium (Na) to uranium (U) between all elements, up to 26 elements simultaneously each time (For use in the precious metals industry only) |
| Analysis Software | qualitative and quantitative analysis software |
| External Dimensions | 460 × 390 × 360 mm |
| Weight | 38kg |
| Instruments Environmental Requirements | |
| Ambient Temperature | 5 °C ~ 38 °C (Suggested that in air conditioning room) |
| Relative Humidity | 15% ~ 90% |
| Power Requirements | AC 220V ± 10V, or AC 110V ± 10V, 50/60 Hz |
| Installation Requirement | No high power electromagnetic and vibration interference sources nearby |
| Analysis Range | ppm to 99.99% |
| Accuracy | 0.03% (Au ≥ 96%) |
| Sample Form | Solid, powder, liquid |
| Tube Voltage | 5kV ~ 50kV |
| High Voltage Power Supply | 0 ~ 50kV |
| Camera | HD camera |
| Filters | Selectable customized switches |
| Detector Type | Si-pin |
| Resolution | 145 ± 5 eV |
| Multichannel Analyzer | Digital multichannel |
| Sample Chamber Size | 310 × 270 × 80 mm |
| Test Time | 5 ~ 60 sec |
| Measuring Element | 82 elements from sodium (Na) to uranium (U) between all elements, up to 26 elements simultaneously each time (For use in the precious metals industry only) |
| Analysis Software | qualitative and quantitative analysis software |
| External Dimensions | 460 × 390 × 360 mm |
| Weight | 38kg |
| Instruments Environmental Requirements | |
| Ambient Temperature | 5 °C ~ 38 °C (Suggested that in air conditioning room) |
| Relative Humidity | 15% ~ 90% |
| Power Requirements | AC 220V ± 10V, or AC 110V ± 10V, 50/60 Hz |
| Installation Requirement | No high power electromagnetic and vibration interference sources nearby |