The AELAB AQMS-200 Dynamic Calibrator utilizes two mass flow controllers to perform standard zero and span calibration with up to four gas sources. It ensures stable trace-level output and supports optional ozone generation, gas phase titration, and integrated photometric measurement for independent ozone calibration.
The system incorporates a high-performance linearized mass flow controller that provides stable and reproducible output for environmental monitoring stations, CEMS platforms, and laboratory calibration facilities.
The AQMS-200 integrates a precision dilution manifold with dual mass flow controllers engineered for high linearity and repeatability. The internal gas pathways are optimized for low-resistance flow distribution, supporting four calibration gas inputs and one diluent gas port. Optional modules include an ozone generator (0.1–6ppm) and a UV photometer for independent ozone measurement. The housing is designed for 220V AC operation with RS232, RS485, and Ethernet interfaces.
The calibrator performs dynamic dilution by regulating flow rates through two electronically controlled mass flow controllers. This enables accurate preparation of target gas concentrations for zero and span calibration. When equipped with the ozone generator, the system produces controlled ozone output with <1%/24h stability. The optional UV photometer quantifies ozone concentration using a precise absorption method, providing ppb-level accuracy, rapid response, and low drift characteristics.
The AQMS-200 supports RS232, RS485, and Ethernet for seamless integration into automated CEMS systems, laboratory calibration benches, and remote environmental networks. Data communication enables real-time monitoring, calibration scheduling, and GLP-compliant reporting.
| Model | Specifications |
|---|---|
| AQMS-200 | 0~10SLPM dilution, 0~100sccm standard gas, optional 0.1~6ppm ozone generator, UV photometer (0~10ppm), RS232/RS485/Ethernet |
The AELAB AQMS-200 Dynamic Calibrator is suitable for ambient air monitoring stations, industrial emissions monitoring, CEMS verification, laboratory reference calibration, and trace-gas validation for environmental and atmospheric research. It supports ozone, NOx, SO₂, CO, and multi-component gas calibration workflows.
The AQMS-200 is engineered under ISO and GLP quality protocols, ensuring calibration traceability, long-term stability, and reproducible performance for regulatory air monitoring and industrial compliance programs.
Specifications :
| Dilution System | |
|---|---|
| Dilution Gas Flow Range | 0~10SLPM, expandable |
| Standard Gas Flow Range | 0~100sccm; expandable |
| Flow Accuracy | ±1% F.S. |
| Linearity | ±(0.5~1)% F.S. |
| Repeatability | ±0.2% F.S. |
| Calibration Gas Input | 4 |
| Diluent Gas Input Ports | 1 |
| Ozone Generator (Optional) | |
| Output | 0.1~6ppm |
| Stability | <1%/24h |
| Linearity | <1% F.S. |
| UV Photometer (Optional) | |
| Range | (0~0.5)ppm, (0~10)ppm |
| Precision | 1.0ppb |
| Linearity | 1% F.S. |
| Rise/Fall Time | <30s |
| Response Time | 180s(90%) |
| Zero Drift | <1.0ppb/24h |
| Range Drift | <1% F.S./24h |
| Physical and Communication | |
| Power Requirement | 220V AC, 50Hz, ≤150W |
| Operating Temperature | 0~40°C |
| Operating Humidity | 0~95%RH (No condensation) |
| Output | RS232/RS485/Ethernet |
| Dimensions and Weight | 609×432×178mm, ≤20kg |
| Dilution System | |
|---|---|
| Dilution Gas Flow Range | 0~10SLPM, expandable |
| Standard Gas Flow Range | 0~100sccm; expandable |
| Flow Accuracy | ±1% F.S. |
| Linearity | ±(0.5~1)% F.S. |
| Repeatability | ±0.2% F.S. |
| Calibration Gas Input | 4 |
| Diluent Gas Input Ports | 1 |
| Ozone Generator (Optional) | |
| Output | 0.1~6ppm |
| Stability | <1%/24h |
| Linearity | <1% F.S. |
| UV Photometer (Optional) | |
| Range | (0~0.5)ppm, (0~10)ppm |
| Precision | 1.0ppb |
| Linearity | 1% F.S. |
| Rise/Fall Time | <30s |
| Response Time | 180s(90%) |
| Zero Drift | <1.0ppb/24h |
| Range Drift | <1% F.S./24h |
| Physical and Communication | |
| Power Requirement | 220V AC, 50Hz, ≤150W |
| Operating Temperature | 0~40°C |
| Operating Humidity | 0~95%RH (No condensation) |
| Output | RS232/RS485/Ethernet |
| Dimensions and Weight | 609×432×178mm, ≤20kg |
Download catalog for:
Download Manual for: