AELAB JS-III Gel Strength Tester for Precise Gelatin Bloom Strength Measurement

The AELAB JS-III Gel Strength Tester provides precise Bloom strength measurement for pharmaceutical, food, and industrial gelatin evaluation. Its intelligent LCD system, regulated water bath, and programmable sampling modes ensure accurate, repeatable gel penetration testing across 5–1000 Bloom. Designed for compliance with Pharmacopoeia standards, the JS-III delivers stable analytical performance for routine and high-precision gel quality assessment.
Modles : JS-III
whatsapp Whatsapp

The AELAB JS-III Gel Strength Tester is mainly used to measure the gel strength of gelatin under stipulated environmental conditions. The gel strength of gelatin is an important indicator to reflect the technological level of gelatin production and control the production quality of gelatin.

The JS-III system includes the host of the gel strength tester, the precise constant temperature monitoring water tank of model HW-III, and the ZL-III efficient refrigerating machine. It meets all requirements for the measurement time limit in the Pharmacopoeia of Edition 2015 and dynamically monitors the press-in depth error of the probe throughout the testing process.

 

⚙️ Design & Construction

The AELAB JS-III Gel Strength Tester is engineered with an intelligent LCD interface, precision displacement system, and regulated water bath structure. The thermostatic bath maintains 5–60 °C with ±0.1 °C accuracy, supported by a magnetic circulating water pump for uniform heat distribution. A high-precision temperature sensor ensures consistent measurement conditions, while the full-automatic frequency-conversion refrigeration system stabilizes the bath rapidly for repeated testing. The integrated cylinder assembly and standardized gel bottles ensure reproducible penetration geometry for high-quality Bloom strength determinations.

 

🔬 Working Principle / Testing Method

The JS-III applies a controlled downward probe movement into a gelatin sample after defined conditioning. The system records penetration depth, force (Bloom g), and stroke displacement with ≤±0.1 mm error. The embedded controller interprets the real-time force curve to determine Bloom strength. Users may select single-step, auto, maintaining, or circular sampling modes. Two-speed penetration modes (normal and high-speed) enable precise testing across a variety of gelatin hardness profiles.

 

💡 Key Advantages

  • Wide 5–1000 Bloom measurement range for pharmaceutical, food, and industrial gelatin testing
  • Programmable sampling depth from 1–60 mm ensures reproducibility across gel types
  • Rapid bath stabilization to 10 °C in under 40 minutes for efficient workflow
  • Multiple sampling speeds (0.1–2 mm/s) for flexible characterization of gel textures
  • Automatic probe depth monitoring eliminates sampling deviation and operator error
  • High-speed measurement mode enhances throughput for batch QC operations
  • Integrated temperature-controlled environment ensures compliance with Pharmacopoeia gel conditioning requirements

 

🔗 Connectivity & Integration

The JS-III supports standard data printing, on-board data storage, and experimental data recording. The system includes automatic fault diagnostics, sensor protection, and non-linear error correction, ensuring GLP-aligned traceability and reliable long-term use.

 

🧠 Model Insight

Model Specifications
JS-III 5–1000 Bloomg range, ±0.3% precision, 1–60 mm sampling depth, 12.700±0.010 mm probe stroke, 60 W power, water bath 5–60 °C ±0.1 °C
HW-III Thermostatic water tank, 58×44×22 cm³ dimensions
ZL-III Refrigerating machine, 48×38×29 cm³ dimensions, (30 °C → 10 °C) < 40 min cooling time

🧪 Testing Applications

The AELAB JS-III is designed for gelatin Bloom strength testing in pharmaceutical capsule production, food-grade gelatin manufacturing, photographic gelatin processing, and industrial colloid quality control. It is widely used in QC laboratories, research institutions, production facilities, and standard compliance testing environments.

 

Compliance & Quality Assurance

The instrument meets the measurement requirements of the Pharmacopoeia of Edition 2015. All AELAB systems follow ISO, CE, and GLP-aligned manufacturing standards, with complete calibration traceability and temperature-control verification.

 

Specifications :

Item Specification
Test range 5–1000 Bloom g
Test precision ±0.3%
Resolution 1 g
Repeated measurement error ±1 Bloom g
Sampling depth 1–60 mm
Stroke error ≤ ±0.1 mm
Stroke error (high-speed measurement) ≤ ±0.1 mm
Measured value take Bloom g (as the unit)
Measured diameter of cylinder 12.700 ± 0.010 mm
Standard gel strength bottle Volume: 150 ml; Inner diameter: 59 mm; Height: 85 mm
Total power 60 W
Overall dimensions  L × W × H 260 mm × 260 mm × 450 mm

Water bath temperature control range 5–60.0 °C
Water bath temperature control precision ±0.1 °C
Refrigerating time (30 °C → 10 °C) < 40 min
Power supply 220 V / 50 Hz / 1200 W
Overall dimensions HW-III (58×44×22) cm³; ZL-III (48×38×29) cm³
Item Specification
Test range 5–1000 Bloom g
Test precision ±0.3%
Resolution 1 g
Repeated measurement error ±1 Bloom g
Sampling depth 1–60 mm
Stroke error ≤ ±0.1 mm
Stroke error (high-speed measurement) ≤ ±0.1 mm
Measured value take Bloom g (as the unit)
Measured diameter of cylinder 12.700 ± 0.010 mm
Standard gel strength bottle Volume: 150 ml; Inner diameter: 59 mm; Height: 85 mm
Total power 60 W
Overall dimensions  L × W × H 260 mm × 260 mm × 450 mm

Water bath temperature control range 5–60.0 °C
Water bath temperature control precision ±0.1 °C
Refrigerating time (30 °C → 10 °C) < 40 min
Power supply 220 V / 50 Hz / 1200 W
Overall dimensions HW-III (58×44×22) cm³; ZL-III (48×38×29) cm³

Frequently Asked Questions

1What is the measurement range of the AELAB JS-III Gel Strength Tester?
The JS-III measures gelatin strength from 5–1000 Bloomg.
2Does the JS-III include temperature-controlled conditioning?
Yes, it uses a 5–60 °C thermostatic bath with ±0.1 °C precision and a rapid refrigeration system.
3What sampling depths can be programmed?
The instrument supports sampling depths from 1–60 mm.
4Can the system operate in high-speed testing mode?
Yes, it provides two penetration speed modes, including high-speed sampling for faster throughput.
5What are the recommended environmental conditions?
The JS-III operates reliably in standard laboratory environments with stable ambient temperature and non-condensing humidity.