The AELAB FKC-1 Microbial Air Sampler performs viable airborne microorganism sampling in cleanrooms, hospitals, food plants, and research facilities. Using impactor-based, equal-flow (isokinetic) sampling at 100 L/min with standard 90 mm Petri dishes, the instrument concentrates microorganisms onto agar for incubation and colony enumeration. AELAB FKC-1 supports contamination control, routine environmental monitoring, and validation workflows in controlled environments.
Modles : FKC - 1
whatsapp Whatsapp

The AELAB FKC-1 Microbial Air Sampler is an impactor-type instrument for viable air monitoring in cleanrooms and controlled environments. The device draws air at a defined flow of 100 L/min and deposits airborne microorganisms onto agar surfaces for incubation and colony enumeration, supporting contamination control and qualification activities across pharmaceutical, healthcare, food, and biotechnology facilities.

⚙ Design & Construction

The AELAB FKC-1 comprises an upper impactor assembly with a perforated sampling head, sampling seat, and pump, and a lower base incorporating the controller and batteries. The Aluminum exterior presents a smooth, non-porous surface that can be sterilized and disinfected prior to sampling. The inlet features multiple precision orifices that distribute the jet pattern, reducing puffball superposition and improving count accuracy on 90 mm Petri dishes.

🔬 Working Principle / Testing Method

The impactor accelerates intake air so that viable particles impinge onto agar in a standard 90 × 15 mm culture dish. The FKC-1 maintains equal-flow (isokinetic) sampling with an inlet impact speed of 0.38 m/s, matching typical cleanroom airflow. After sampling a defined volume (0.001–9.999 m³), plates are incubated to obtain CFU counts, quantifying microbial concentration as CFU per cubic meter of air.

💡 Key Advantages

  • Isokinetic accuracy: AELAB FKC-1 regulates inlet velocity at 0.38 m/s to reflect true cleanroom airflow during viable air sampling.
  • High sampling throughput: Fixed 100 L/min flow supports time-efficient monitoring during qualification and routine EM rounds.
  • Standardized consumables: Uses common 90 mm agar plates for broad media compatibility and validated recovery.
  • Count integrity: Multi-hole impactor head minimizes colony coalescence for more reliable CFU enumeration.
  • Portable operation: DC 16.8 V rechargeable system provides approximately 4 h runtime for field or room-to-room surveys.

🔗 Connectivity & Integration

The AELAB FKC-1 integrates into microbiological environmental monitoring by pairing with incubators and colony counting workflows. Results are recorded as sampled volume and subsequent CFU counts on the plate. The instrument is compatible with SOP-driven EM programs and trending in LIMS or quality systems using manual data entry from plate counts.

🧠 Model Insight

Model FKC-1 centers on equal-flow impact sampling for viable particle capture. The perforated head geometry and controlled 100 L/min flow stabilize jet impaction across the 90 mm agar surface. The instrument’s cylindrical form factor (φ120 × 325 mm; 4.0 kg) and Aluminum build facilitate handheld or tripod-based use in clean zones.

🧪 Testing Applications

  • Pharmaceutical cleanroom viable air monitoring during qualification and routine EM.
  • Hospital sterile area surveillance for airborne pathogens.
  • Food and beverage QC area monitoring for hygienic zoning.
  • Biotechnology R&D spaces and small-scale production suites.

Compliance & Quality Assurance

The AELAB FKC-1 supports viable air sampling methods commonly used in cleanroom environmental monitoring programs. Use according to site SOPs and applicable regulatory guidance; verify calibration and media recovery per quality procedures.

Specifications :

Feature Details
Capacity Sampling flow 100 L/min
Readability Sampling volume settable 0.001–9.999 m³
Units m³ sampled; CFU after incubation on agar
Platform Size 90 × 15 mm Petri dish (φ90 × 15 mm)
Power DC 16.8 V rechargeable battery
Battery Approx. 4 h operation per charge
Calibration Isokinetic inlet impact speed 0.38 m/s
Dimensions / Weight φ120 × 325 mm / 4.0 kg
Construction Aluminum housing; perforated sampling inlet to minimize colony overlap
Operating Conditions 10–35 °C; 10–90 % RH; 80–110 kPa
Accessories Rechargeable battery (included); Petri dish holder (90 mm)
Feature Details
Capacity Sampling flow 100 L/min
Readability Sampling volume settable 0.001–9.999 m³
Units m³ sampled; CFU after incubation on agar
Platform Size 90 × 15 mm Petri dish (φ90 × 15 mm)
Power DC 16.8 V rechargeable battery
Battery Approx. 4 h operation per charge
Calibration Isokinetic inlet impact speed 0.38 m/s
Dimensions / Weight φ120 × 325 mm / 4.0 kg
Construction Aluminum housing; perforated sampling inlet to minimize colony overlap
Operating Conditions 10–35 °C; 10–90 % RH; 80–110 kPa
Accessories Rechargeable battery (included); Petri dish holder (90 mm)

Product Video:

Frequently Asked Questions

1What testing applications is the AELAB FKC-1 designed for?
Viable air monitoring in pharmaceutical cleanrooms, hospital sterile rooms, food processing areas, biotechnology labs, and controlled research facilities.
2What sampling flow and volume settings does the FKC-1 support?
The sampler operates at 100 L/min with a user-settable sampling volume from 0.001 to 9.999 m³.
3Which culture plates does the instrument use?
Standard 90 × 15 mm (φ90 × 15 mm) Petri dishes compatible with common agar media.
4How does the AELAB FKC-1 ensure accurate results?
The impactor maintains isokinetic sampling with an inlet speed of 0.38 m/s and uses a multi-hole head to reduce colony overlap.
5What are the operating conditions?
10–35 °C, 10–90 % RH, and 80–110 kPa.
6How is the unit powered and how long does it run?
A DC 16.8 V rechargeable battery provides approximately 4 hours of operation per charge.
7Does the FKC-1 provide data outputs or ports?
Specifications provided focus on sampling flow and mechanics. Confirm interfaces and data export options with AELAB if needed.

Looking for specific lab equipment? Fill out the form below, and our team will get back to you with detailed information and a personalized quote.

Request Quote: Cleanroom Air Sampler FKC-1 – AELAB