info@aelabgroup.com
QR Code 1
QR Code 2

AELAB MicroPette Plus 8-Channel Adjustable Volume Pipette (magnetic suction model) 7041 Series

The AELAB MicroPette Plus 8-Channel Adjustable Volume Mechanical Pipette is engineered for precision and user comfort. With ergonomic design, chemical resistance, and compatibility with 96-well plates, it ensures reproducible results in high-throughput laboratories.
Modles : 7041 Series
whatsapp Whatsapp

The AELAB MicroPette Plus 8-Channel Adjustable Pipette is engineered for laboratories requiring reliable, precise, and efficient liquid handling. Featuring a lightweight ergonomic design, smooth plunger action, and compatibility with standard microplates, this pipette enhances productivity and ensures reproducibility in research, diagnostics, and quality control labs.

 

🔬 How It Works?

The AELAB MicroPette Plus uses an air-displacement mechanism to aspirate and dispense liquids with precision. Its rotating counter knob and digital display allow quick and accurate volume adjustment, while the 8-channel format is optimized for 96-well plate applications. Autoclavable and UV-resistant materials ensure durability and contamination-free use.

 

💡 Advantages

  • Ergonomic Design: Light-touch plunger and finger rest reduce hand fatigue

  • Versatility: Available in multiple volume ranges for diverse applications

  • High Throughput: 8-channel design fits standard 96-well plates

  • Durability: Resistant to chemicals, heat, and UV sterilization

  • Easy Maintenance: Fully autoclavable with replaceable parts and online calibration

  • Certified Quality: ISO 9001:2008, ISO 13485:2003, and CE marked

 

🔍 Recommended Models by Use Case

  • Educational & Teaching Labs: 0.5–10 μL and 5–50 μL ranges for micro-volume handling

  • Research Centers: 10–100 μL model for DNA, RNA, and protein studies

  • Quality Control & Diagnostics: 50–300 μL for medium-volume assays

  • High-Throughput Screening: 100–1000 μL model for ELISA, drug testing, and clinical applications

 

🧪 Ideal For

  • Molecular biology & genomics labs

  • Pharmaceutical & biotech R&D

  • Food and chemical quality control labs

  • Clinical diagnostics and hospital labs

  • High-throughput research requiring 96-well plates

Specifications :

Cat. No. Volume Range Increment Test Volume Systematic Error Random Error Channels
7041303004 0.5–10 μL 0.1 μL 10 μL / 5 μL / 1 μL ±0.15 μL ±1.50% / ±0.125 μL ±2.50% / ±0.04 μL ±4.00% ±0.15 μL ±1.50% / ±0.125 μL ±2.50% / ±0.04 μL ±4.00% 8
7041303006 5–50 μL 0.5 μL 50 μL / 25 μL / 5 μL ±0.50 μL ±1.00% / ±0.375 μL ±1.50% / ±0.15 μL ±3.00% ±0.25 μL ±0.50% / ±0.25 μL ±1.00% / ±0.10 μL ±2.00% 8
7041303008 10–100 μL 1 μL 100 μL / 50 μL / 10 μL ±0.8 μL ±0.80% / ±0.5 μL ±1.00% / ±0.3 μL ±3.00% ±0.3 μL ±0.30% / ±0.4 μL ±0.80% / ±0.2 μL ±2.00% 8
7041303012 50–300 μL 5 μL 300 μL / 150 μL / 50 μL ±2.10 μL ±0.70% / ±1.50 μL ±1.00% / ±0.75 μL ±1.50% ±0.75 μL ±0.25% / ±0.75 μL ±0.50% / ±0.40 μL ±0.80% 8
7041303014 100–1000 μL 10 μL 1000 μL / 500 μL / 100 μL ±10 μL ±1.00% / ±10 μL ±2.00% / ±5 μL ±5.00% ±3 μL ±0.30% / ±2.5 μL ±0.50% / ±1 μL ±1.00% 8
Cat. No. Volume Range Increment Test Volume Systematic Error Random Error Channels
7041303004 0.5–10 μL 0.1 μL 10 μL / 5 μL / 1 μL ±0.15 μL ±1.50% / ±0.125 μL ±2.50% / ±0.04 μL ±4.00% ±0.15 μL ±1.50% / ±0.125 μL ±2.50% / ±0.04 μL ±4.00% 8
7041303006 5–50 μL 0.5 μL 50 μL / 25 μL / 5 μL ±0.50 μL ±1.00% / ±0.375 μL ±1.50% / ±0.15 μL ±3.00% ±0.25 μL ±0.50% / ±0.25 μL ±1.00% / ±0.10 μL ±2.00% 8
7041303008 10–100 μL 1 μL 100 μL / 50 μL / 10 μL ±0.8 μL ±0.80% / ±0.5 μL ±1.00% / ±0.3 μL ±3.00% ±0.3 μL ±0.30% / ±0.4 μL ±0.80% / ±0.2 μL ±2.00% 8
7041303012 50–300 μL 5 μL 300 μL / 150 μL / 50 μL ±2.10 μL ±0.70% / ±1.50 μL ±1.00% / ±0.75 μL ±1.50% ±0.75 μL ±0.25% / ±0.75 μL ±0.50% / ±0.40 μL ±0.80% 8
7041303014 100–1000 μL 10 μL 1000 μL / 500 μL / 100 μL ±10 μL ±1.00% / ±10 μL ±2.00% / ±5 μL ±5.00% ±3 μL ±0.30% / ±2.5 μL ±0.50% / ±1 μL ±1.00% 8

No download files available for this product.

Frequently Asked Questions

1How does the AELAB MicroPette Plus maintain accuracy across multiple channels?
Each channel is calibrated individually with precision-engineered pistons and seals, ensuring uniform volume delivery across all 8 tips.
2Can the pipette be sterilized without disassembly?
Yes, the entire unit is autoclavable or UV-disinfectable without dismantling, maintaining sterility with minimal effort.
3What makes the MicroPette Plus suitable for 96-well plates?
The 8-channel design aligns perfectly with standard microplate rows, enabling efficient parallel pipetting for assays and screenings.
4How is calibration managed?
Calibration can be done using standard lab tools or via the AELAB online calibration platform for quick, user-friendly adjustments.
5What safety certifications does it have?
The pipette is ISO 9001:2008, ISO 13485:2003 certified, and CE marked, ensuring compliance with international quality and safety standards.

Looking for specific lab equipment? Fill out the form below, and our team will get back to you with detailed information and a personalized quote.

Request Quote: AELAB MicroPette Plus 8-Channel Adjustable Volume Pipette (magnetic suction model) 7041 Series